Kvantovaya Elektronika, 1992 Volume 19, Number 3, Pages 274–276 (Mi qe3393)

This article is cited in 1 paper

Control and measurement of laser radiation parameters

Comparative measurements of the duration of picosecond light pulses by direct and autocorrelation techniques

E. G. Pivinskii, G. E. Nikolaev, D. S. Prilezhaev, Z. V. Nesterova


Abstract: Light pulse durations in the range 20–60 ps were measured by the simultaneous use of a direct method (with an Agat SF-1 image-converter camera) and a correlation technique (noncollinear second harmonic generation in a single pulse). The diameter of the fundamental radiation beam was shown to influence the results of measuring the duration of light pulses by the correlation technique. Practical recommendations for correct measurements by this technique were developed.

UDC: 621.373.826

PACS: 42.65.Re, 42.79.Ls, 42.65.Ky, 42.30.-d

Received: 26.07.1991


 English version:
Soviet Journal of Quantum Electronics, 1992, 22:3, 250–252