A Collison nebulizer as an ion source for mass spectrometry analysis

V. V. Pervukhin, D. G. Sheven, Yu. N. Kolomiets

Nikolaev Institute of Inorganic Chemistry, Siberian Branch of the Russian Academy of Sciences, Novosibirsk

Abstract: It is proposed to use a Collison nebulizer as a source of ionization for mass-spectrometry with ionization at atmospheric pressure. This source does not require an electric voltage, radioactive sources, heaters, or liquid pumps. It is shown that the number of ions produced by the Collison nebulizer is ten times greater than the quantity of ions produced by the $^{63}$Ni radioactive source and three to four times greater than the number of ions produced with sonic ionization devices.